What HPLC measures
High-performance liquid chromatography (HPLC) separates a mixture into its components by pushing it through a packed column at high pressure. For research peptides, reversed-phase HPLC on a C18 column is the workhorse: molecules interact with a hydrophobic stationary phase and elute in order of increasing hydrophobicity as the mobile phase becomes more organic.
A UV detector at 214 nm records the absorbance of the peptide bond as material leaves the column. The area under each peak in the resulting chromatogram is proportional to the amount of that species. Purity is reported as the target peak's area divided by the total area of all detected peaks.
A defensible HPLC method
A method that supports a purity claim discloses at minimum:
- Column: chemistry (usually C18), particle size, pore size, length, internal diameter.
- Mobile phases: typically A = 0.1% TFA in water, B = 0.1% TFA in acetonitrile.
- Gradient: starting and ending %B, total run time, hold times.
- Flow rate: usually 0.2 to 1.5 mL/min depending on column geometry.
- Column temperature.
- Injection volume and sample concentration.
- Detection wavelength: 214 nm is standard for peptide bond detection.
A number without a method is unverifiable. A conservative gradient — for example, 5 to 65 percent B over 20 to 30 minutes — resolves close-eluting impurities that a fast gradient will hide.
Reading the chromatogram
- The main peak should be dominant, symmetric, and cleanly baseline-separated from anything nearby.
- Impurity peaks should be small, and — for reproducible material — appear at consistent retention times lot to lot.
- Baseline noise should be flat before the first peak and after the last, with no unexplained rises.
- A peak-integration table should list retention time and percent area for each detected peak, not just the target.
What HPLC does not tell you
HPLC purity is a measure of chromatographic homogeneity under a specific method. It does not confirm identity — that requires mass spectrometry. It does not measure endotoxin, bioburden, moisture, or counter-ion content. And it says nothing about the integrity of the vial after it leaves the lab. A complete lot release combines HPLC with LC-MS and — where relevant — sterility and endotoxin assays.
For the identity piece, continue with How to Read a Peptide COA or What Third-Party Testing Means. For sterility and endotoxin, see Sterility and Endotoxin Testing.
